AeromechTM Modular Biosafety Cabinet Type A1 & A2 are designed for the security and protection of operatorââ¬â¢s product and environment from pathogenic agents and microorganisms transmitted in respiratory way and also for creation of a bacterial dust free air environment. It has a common plenum from which 30% of air is exhausted, and 70% re-circulated to the work area as part of the downflow. Type A cabinets drain out air directly back to the laboratory, and they may contain positive pressure contaminated plenums. When toxic chemicals must be employed as an adjunct to microbiological processes, these cabinets should not be used. Exhaust HEPA filtration only.
Removes airborne aerosols including biohazards, and not chemical fumes.
The Class II Type A1 consists of the positively-pressurized contaminated plenum bordering the ambient environment, and therefore is less safer than the Class II Type A2 that has a negative pressure surrounding the positively pressurized contaminated plenum. In any unlikely event of a leakage on the positive plenum, the leaking aerosol will be pulled by the negative pressure back to the positive plenum, which prevents it from leaking out. On the grounds of safety, the Type A1 design is now considered obsolete. In the A2 cabinet, approximately 70% of air from the positive plenum is re-circulated as downflow.
|
Product
|
Class II Bio safety Cabinet-Type A1
|
|
Air balancing
|
70% recirculation, 30% exhaust
|
|
Protection
|
Product & personnel & environmental protection
|
|
Bio safety Level
|
Bio safety Level 1, 2 & 3
|
|
Exhaust Duct Requirement
|
Inside the room / Thimble duct
|
|
Control Plenum surrounded by
|
Outside Air
|
|
Cleanliness level
|
ISO class 5 as per ISO 14644-1
|
|
Average Airflow Velocity
|
Inflow Velocity
|
Minimum 0.45 m/s ( 90 FPM) at 8" front opening
|
|
Down Velocity
|
0.30 m/s ( 65 FPM)
|
|
Noise level
|
60 db on A scale
|
|
Light Intensity
|
Minimum 900 lux inside the working area at ambient zero
|
|
MOC
|
Main Body
|
Galvanized Iron with PU coated / Stainless Steel - Satin Finish
|
|
Working Area
|
Stainless Steel IS 304 Grade Satin Finish
|
|
Supply Filter
|
HEPA Filter having EU 13 rating with suitable size.
|
|
Efficiency
|
99.99% down to 0.3 Microns.
|
|
Exhaust Filter
|
HEPA Filter having EU 13 rating with suitable size.
|
|
Efficiency
|
99.99% down to 0.3 Microns.
|
|
Pre Filter
|
Aeromechî make, EU 4 rating with suitable size.
|
|
Efficiency
|
90% down to 10 Microns.
|
|
Noise Level
|
Less than 70 db 'A'.
|
|
Vibration Level
|
Minimum.
|
|
Instrumentation |
|
Differential Pressure Gauge
|
Magnehelic Gauge - 01 No. (make : Dwyer, USA) with each unit (To measure differential pressure across the supply HEPA filter).
|
|
Motor Blower Assembly
|
Statically & Dynamically balanced with suitable rating and size. (Induction motor with centrifugal blower) A Mounting unique spring suspension system to take care of vibration and noise.
|
|
3 Pins Switch & Socket
|
15 AMPS - 1 set in each unit.
|
|
Other Switches
|
On/Off switches for motor blower & CFL lights.
|
|
Microprocessor Control
|
Airflow display and UV Interlocking and Low airflow Alarm and Supply & Exhaust filter Fail
|
|