We offer Operation Theater Version of Vertical Laminar Airflow systems. These are used in hospitals for maintaining control over potential environmental contamination. These are available in standard specifications and at very competent prices.
|
Product
|
Aeromechî Vertical laminar Airflow -ââ‰â¬Å OT Version
|
|
Type
|
Re circulatory type, Standard type
|
|
Dimensions
|
As per drawings enclosed.
|
|
MOC
|
Galvanized Iron with PU coated / Stainless Steel - Satin Finish / Wooden laminated
|
|
Pre Filter
|
Aeromech Make Pre Filter with Micro Glass Fibre Media of EU - 4 Rating.
|
|
Supply HEPA Filter
|
Imported Minipleat HEPA Filter efficiency 99.97% Down to 0.3õ H-13 Rating.
|
|
Motor Blower Assembly
|
Induction Motor with Centrifugal Blower Statically and Dynamically balanced with Suitable Rating and Size.
|
|
Mounting
|
Unique Spring Suspension Systems to take care of Vibration and Noise.
|
|
Working Table
|
S.S. 304 with Stabilised Heating Table with Microscope Arrangement
|
|
Side Panel Accessory No.
|
Clear Glass Side Panels/ Clear Acrylic panel/ UV Protected Polycarbonate panel
|
|
Pressure Gauge
|
DWYER, USA Minihelic Differential Pressure Gauge/ Aeromech Inclined manometer.
|
|
Lights
|
Fluorescent Tube with Milky White Diffuser.
|
|
Other Electricals
|
3 Pin Socket, 15 AMP - 1 No. - On / Off Switches with Indicator.
|
|
Performance
|
|
Air Cleanliness
|
Class ISO - 5 (as per ISO - 14644-1).
|
|
Air Flow Rate
|
0.45 + 0.05mps. or 90 FPM + 10 FPM
|
|
Illumination Of Light
|
Minimum 800 Lux on work table
|
|
Noise Level
|
Less than 67 db on Scale 'A' When Ambient sound Level is not Greater than 57 db on Scale 'A'
|
|
Vibration Level
|
Minimum.
|
|